Loading Quick Save Presets from Removable Media
When you load setups into the system from removable media, the contents of
the source storage media are displayed; however, you cannot select specific
items to load. When you load setups from removable media, all setups on that
storage media are loaded onto the system. To load specific setups, copy only
the required setups to removable media and then load them to the new location.
1. Press the Setup key.
2. In the setups, click Service, and then click System Management.
3. Click Backup/Restore.
4. Insert the removable media containing the settings into the drive.
5. From the Select the Media menu, select the media type from which you
are loading settings.
6. Click Restore.
System Setups
Setups are system parameters that you can change. By changing setups, you can
customize the system to meet your operating preferences. Setups are organized
into seven standard categories: System Settings, Analysis Config
(configuration), Annotate, Print/Network, Audio/Video, Protocols, and
Service. Additionally, two other categories appear only if the corresponding
options are installed on the system. These optional categories are Voice Control
and Data Security. About is not actually a setup category; it is used to display
information about the system software. Changes in setups take effect immediately
and remain in effect until you change them again or load setups from a DVD.
NOTE
The institution name exported with DICOM data always reflects the name shown
in the setups at the time the study ended. When you change the institution name
in the setups, Philips recommends restarting the system after changing the
institution name.
175
iU22 User Manual
4535 614 45861
6
Customizing the System