Disinfectants Compatibility
Read this information before performing disinfection and sterilization procedures.
It discusses recommended disinfectants and choosing an appropriate disinfectant
for the required level of disinfection. You must check "Disinfectants Compatibility
Table" on page 304 for the chemical compatibility of disinfectants and cleaners
with specific transducers. In addition, the table indicates if a device can be sprayed
or wiped only, or if it can be soaked.
WARNINGS
• Not all disinfectants are effective against all types of contamination. Ensure
the disinfectant type is appropriate for the type of transducer and that the
solution strength and time of contact are appropriate for the intended
clinical use.
• Disinfectants listed in this section are recommended because of their
chemical compatibility with product materials, not their biological
effectiveness. For the biological effectiveness of a disinfectant, see the
guidelines and recommendations of the disinfectant manufacturer, the
Association for Professionals in Infection Control, the U.S. Food and Drug
Administration, and the U.S. Centers for Disease Control.
• If a premixed solution is used, be sure to observe the solution expiration
date.
• Always use protective eyewear and gloves when cleaning and disinfecting
any equipment.
CAUTION
Using a non-recommended disinfection solution, using an incorrect solution
strength, or immersing a transducer deeper or longer than recommended can
damage the device and voids the warranty.
iU22 User Manual
302
4535 614 45861
Transducer Care
13