• Alternatively, you may gently wipe the distal tip and flexible shaft up to the
control housing (steering mechanism) with gauze pads soaked in mild,
soapy water. Do not use iodine-based soaps.
3. Use water to rinse the distal tip and flexible shaft thoroughly.
CAUTIONS
• Do not bend the shaft into a circle with a diameter of less than 0.30 m (1
ft).
• Do not use bleach on the transducer and shaft.
• Do not use isopropyl alcohol-based products on the transducer and shaft.
• Do not soak the transducer for extended periods of time. Limit the time
that transducers are soaked in disinfectant solution to the minimum time
recommended by the disinfectant manufacturer.
• Do not rinse or immerse the connector or the portion of the cable near
the connector.
• Do not immerse or rinse the steering mechanism.
• Follow the recommendations of the disinfectant manufacturer.
4. Disinfect the distal tip and flexible shaft by placing them in the appropriate
disinfectant.
5. Remove the tip and shaft from the disinfectant and thoroughly rinse with
water according to the instructions from the disinfectant manufacturer.
6. Check the transducer for any residual organic material. If any is present,
remove it and disinfect the transducer again.
7. Gently dry the distal tip and flexible shaft with a sterile cloth or pad, or allow
it to air dry.
CAUTION
The transducer steering mechanism is not sealed. If disinfectant or other fluid
enters the steering mechanism, it will corrode the gears and electrical
connections. Avoidable transducer damage is not covered by the warranty
or service contract.
297
iU22 User Manual
4535 614 45861
13
Transducer Care