Restricted Use of Isopropyl Alcohol to Clean Transducers
Cable1
Strain relief2
Strain relief/housing joint3
Housing4
2.5 cm (1 inch)5
You can use alcohol in this area6
Do not use alcohol in this area7
1. After cleaning the transducer and cable (see "General Cleaning for All
Transducers" on page 287), wipe or spray the transducer and cable with a
low-level disinfectant. Allow for the manufacturer’s recommended contact
time. For a list of compatible disinfectants, see "Disinfectants Compatibility"
on page 302.