WARNING
Before performing maintenance or cleaning, always press the On/Off switch to
turn the system off, set the power switch to off, and then disconnect the system
from the wall outlet.
CAUTIONS
Attempting to disinfect a cable or connector by using a method other than
the one included here can damage the device and voids the warranty.
Orient the parts that must remain dry higher than the wet parts until all
parts are dry.
Use the following procedure to clean the display; the system control panel; and
all the external surfaces of the system and the cart; the ECG trunk cables, leads,
and electrodes.
1. Before cleaning, turn off the system and unplug the power cord from the
power source.
2. Wipe with a soft cloth moistened with soap and water.
CAUTION
Do not spill or spray liquid on the controls, into the system cabinet, or into
the transducer connection receptacle.
3. Remove any solid matter around the keys or the controls with a cotton swab
or toothpick to ensure that solids are not pushed into the cabinet.
4. If blood or other infectious material comes in contact with the system or any
cable other than a transducer cable, wipe with a 70% solution of isopropyl
alcohol.
WARNING
Before performing maintenance or cleaning, always press the On/Off switch
to turn the system off, set the power switch to off, and then disconnect the
system from the wall outlet.
iU22 User Manual
312
4535 614 45861