Philips iU22 - Output Signals; Data Connections; Modality Interface; Physio

Philips iU22
342 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
• High- and low-level ECG
• Physio pulse, phono, auxiliary 1, and auxiliary 2
• DVR and VCR audio (left and right)
• Microphone for DVR and VCR voice recording
• External peripheral device remote control
Output Signals
• Video: composite color, interlaced black-and-white, S-Video, and DVI-D
• VCR audio (left and right)
• External printer USB serial data
• Physio analog signal
Data Connections
• Ethernet network (10/100/1000 Mb/s)
• USB 2.0
Modality Interface
DICOM standard. DICOM conformance statements for Philips products are on
this Web site:
www.healthcare.philips.com/us/about/connectivity/dicom_conformance_main.wpd
Physio
• Lower Frequency Cut-off: 0.70 Hz ±10%
• Upper Frequency Cut-off: 17 Hz ±10%
• Nominal Input Amplitude: ±5 mV peak
• Minimum QRS Wave Amplitude: 0.05 mV
Peripherals
• S-VHS VCR
• DVR
• B/W video printer
• Color video printer
• Report printer
iU22 User Manual
328
4535 614 45861
Specifications