other electrical devices used within the same room or an adjacent room, or
from portable and mobile RF communications equipment such as cellular phones
and pagers, or from the existence of radio, TV, or microwave transmission
equipment located nearby. In cases where electromagnetic interference (EMI)
is causing disturbances, it may be necessary to relocate your system.
The system complies with International Standard CISPR 11 for radiated and
conducted electromagnetic disturbances. Compliance with this standard allows
the system to be used in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
WARNING
Using cables, transducers, or accessories other than those specified for use with
the system may result in increased emissions or decreased immunity of the
system.
CAUTION
Medical equipment has special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the
system’s accompanying documents.
This section includes information on electromagnetic emissions and immunity
as it applies to the system. Ensure that the operating environment of your system
meets the conditions specified in the referenced information. Operating the
system in an environment that does not meet these conditions may degrade
system performance.
The information and warnings contained in this and other sections should be
observed when installing and using the system to ensure its EMC.
NOTE
See the other electrical-safety warnings and cautions in this section.