Electromagnetic Immunity: Environment Guidance
Electromagnetic
Environment Guidance
Compliance
Level
IEC 60601
Test Level
Immunity
Test
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Same as
IEC 60601 test
level
± 6 kV contact,
± 8 kV air
Electrostatic
discharge
(ESD), IEC
61000-4-2
Mains power quality should be
that of a typical commercial or
hospital environment.
Same as
IEC 60601 test
level
± 2 kV for
power supply
lines, ± 8-18 kV
for
input/output
lines
Electrical fast
transient/burst,
IEC
61000-4-4
Mains power quality should be
that of a typical commercial or
hospital environment.
Same as
IEC 60601 test
level
± 1 kV
differential
mode, ±2 kV
common mode
Surge, IEC
61000-4-5