5 Managing Patients and Equipment
102
PIIC iX With the IntelliVue Information Center iX, the following additional measurement settings can be
synchronized:
WARNING
Not all settings are synchronized; after changing the measurement source, always check that the
settings are appropriate.
If later the patient is disconnected from the other device, and reconnected to the monitor again, any
changes in the settings made in the meantime will be passed on to the monitor. In this way, settings
continuity is preserved when the measurement source changes.
NOTE
Settings synchronization can be switched off at the monitor in Configuration mode.
In certain situations, you will be asked to confirm that synchronization of settings is appropriate.
Settings synchronization can only take place when there is no patient information mismatch between
the monitor and the Information Center.
PIIC If a Check ECG Settings or !! Check Pairing INOP appears always check that the ECG settings, especially
the paced setting, are appropriate for your patient.
PIIC iX If a Check ECG Settings, Chk SpOT Settings, or another check settings INOP appears, always check that
the settings, especially the paced setting for ECG, are appropriate for your patient.
ST ST Analysis On/Off, ST Alarm On/Off, ISO Point, J point, ST point, ST
Alarm Limits, ST-Index On/Off, ISO/J-Point Detection, ST Lead On/
Off, ST Baseline
QT QT analysis On/Off, QT Lead, QTc High Alarm On/Off, ΔQTc Alarm
On/Off, QTc High Limit, ΔQTc High Limit, QT Baseline
2
SpO
2
TSpO
2
Alarms on/off, SpO
2
Alarm limits, SpO
2
Low Alarm Delay, SpO
2
High Alarm Limit, Desat Alarm Limit
NBP Alarm Suppression On/Off, Pulse(SpO
2
) On/Off
1
ECG On/Off setting is forced to ON by the monitor, if the local setting is ON.
2
With PIIC, QT baseline is synchronized, but not QT snippets.
ECG/Arrhythmia Some ECG Alarms INOP On/Off, Hexad Reconstruction Mode
ST STE On/Off, STE Alarm On/Off
NBP Sys/Dia/Mean Alarm Limits, Alarms On/ Off, Alarm Source
Resp Apnea Time, Alarm Limits, Alarm On/Off, Resp On/Off
SpO
2
SpO
2
Alarms on/off, SpO
2
Alarm limits, Desat Alarm Limit, NBP Alarm
Suppression On/Off, Pulse(SpO
2
) On/Off, Measurement Mode
1
,
Repetition Time
1
1
Measurement Mode and Repetition Time can only be synchronized if SpO
2
comes from a Cableless SpO
2
Pod or a
telemetry device.