24 Specifications
298
Measurement Specifications
See the “Default Settings Appendix” for a list of the settings the monitor is initially shipped with.
ECG/Arrhythmia/ST/QT
Complies with:
• IEC 60601-2-25:1993 + A1:1999/EN60601-2-25:1995 + A1:1999
• IEC 60601-2- 27:2005/EN60601-2-27:2006
• IEC 60601-2-51:2003/EN 60601-2-51:2003
• AAMI EC11/EC13:1991/2002
ECG/Arrhythmia/ST Performance Specifications
Cardiotach Range Adult/pedi: 15 to 300 bpm
Neo range: 15 to 350 bpm
Note: for rates equal to or less than 15 bpm, the displayed heart
rate is 0.
Accuracy ±1% of range
Resolution 1 bpm
Sensitivity ≥200 µV
peak
PVC Rate Range 0 to 300 bpm
Resolution 1 bpm
ST Numeric Range -20 to +20 mm
Accuracy ±0.5 mm or 15%, whichever is greater
Resolution 0.1 mm
QT Numeric Range 200 to 800 ms
Accuracy ±30 ms
Resolution 8 ms
QTc Numeric Range 200 to 800 ms
Resolution 1 ms
ΔQTc Numeric Range -600 to +600 ms
Resolution 1 ms
QT-HR Numeric Range - adult 15 to 150 bpm
Range - pediatric and neonatal 15 to 180 bpm
Sinus and SV Rhythm Ranges Brady Adult: 15 to 59 bpm
Pedi: 15 to 79 bpm
Neo: 15 to 89 bpm
Normal Adult: 60 to 100 bpm
Pedi: 80 to 160 bpm
Neo: 90 to 180 bpm
Tachy Adult: >100 bpm
Pedi: >160 bpm
Neo: >180 bpm