25 Default Settings Appendix
323
Temperature Default Settings
Invasive Pressure Default Settings
Low Limit 90/50 (60) 70/40 (50) 40/20 (24)
Alarms On On On
NBP On On On
Repeat Time 10 min 10 min 10 min
Pulse (NBP) On On On
Unit mmHg mmHg mmHg
Done Tone Off Off Off
Start Time Synchronized Synchronized Synchronized
VP Pressure
60 mmHg 40 mmHg 30 mmHg
Reference Auscultatory Auscultatory Invasive
Color Red Red Red
NBP Settings
Factory Defaults
Adult Pedi Neo
Temp Settings
Factory Defaults
Adult Pedi Neo
Low Limit
36 36 36
High Limit 39 39 39
Alarms On On On
Unit °C °C °C
Range 35...43 35...43 35...43
Color Green Green Green
ABP, ART, Ao, BAP, FAPP,PP1, P2, P3, P4, UAP
Settings
Factory Defaults
Adult Pedi Neo
Alarms from Sys. Sys. Sys.
High Limit
160/90 (110) 120/70 (90) 90/60 (70)
Low Limit 90/50 (70) 70/40 (50) 55/20 (36)
Alarms On On On
Extreme Alarms Disabled Disabled Disabled
Δ Extreme High
15 10 5
Δ Extreme Low 15 10 5
Sys. High Clamp 190/100 (125) 140/80 (100) 105/75 (75)
Sys. Low Clamp 80/45 (65) 60/35 (45) 45/15 (30)
Scale 150 100 100
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz