22 Maintenance and Troubleshooting
258
Inspecting the Cables and Cords
1 Examine all system cables, the power plug and cord for damage. Make sure that the prongs of the
plug do not move in the casing. If damaged, replace it with an appropriate Philips power cord.
2 Inspect the patient cables, leads and their strain reliefs for general condition. Make sure there are
no breaks in the insulation. Make sure that the connectors are properly engaged at each end to
prevent rotation or other strain.
3 Apply the transducer or electrodes to the patient, and with the monitor switched on, flex the
patient cables near each end to make sure that there are no intermittent faults.
Maintenance Task and Test Schedule
The following tasks are for Philips-qualified service professionals only. All maintenance tasks and
performance tests are documented in detail in the service documentation supplied on the monitor
documentation DVD.
Ensure that these tasks are carried out as indicated by the monitor's maintenance schedule, or as
specified by local laws. Contact a Philips-qualified service provider if your monitor needs a safety or
performance test. Clean and disinfect equipment to decontaminate it before testing or maintaining it.
Troubleshooting
If you suspect a problem with an individual measurement, read the Instructions for Use and double-
check that you have set up the measurement correctly.
If you suspect an intermittent, system-wide problem call your service personnel. You may be asked for
information from the status log. To view the status log,
In the
Main Setup menu, select Revisions.
1 Select Status Log.
Type Maintenance and Test Activities Frequency
Monitor Tests Safety checks. Selected tests on the basis of
IEC 60601-1.
At least once every two years, or as needed, after
any repairs where the power supply is removed
or replaced, or if the monitor has been dropped.
Monitor
Maintenance
Check ECG synchronization of the monitor
and defibrillator (only if hospital protocol
requires use of monitor during defibrillation).
At least once every two years, or as needed.
Measurement Tests Performance assurance for all measurements
not listed below.
At least once every two years, or if you suspect
the measurement values are incorrect.
Measurement
Maintenance
NBP calibration At least once every two years, or as specified by
local laws.
Microstream CO
2
calibration and
performance test
At least once a year or after 4000 operating
hours.
Mainstream and sidestream CO
2
calibration
and sidestream flow check
At least once a year, or if you suspect the
measurement values are incorrect.
Battery Maintenance See “Optimizing Battery Performance” on page 254 for battery maintenance activities.