11 Monitoring NBP
189
Enabling Sequence Mode and Setting Up The
Sequence
1 In the Setup NBP menu, select Mode and select Sequence from the pop-up menu.
2 Select Setup Sequence to open the Setup Sequence window.
Up to four measurement cycles can be set up which will run consecutively. For each cycle you can
set the number of measurements and the interval between them. If you want to run less than four
cycles in a sequence, you can set the number of measurements for one or more cycles to
Off.
3 Select each sequence in turn and select the number of measurements and the time interval between
the measurements.
4 To have measurements continue after the sequence, set the number of measurements for your last
cycle to
Continuous and this cycle will run indefinitely.
CAUTION
Be aware that, if none of the cycles are set to Continuous, NBP monitoring will end after the last
measurement of the cycle.
When the NBP measurement mode is set to Sequence, the repetition time for Auto mode cannot be
changed.
Choosing the NBP Alarm Source
You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly or in
parallel. Only one alarm is given, with the priority of mean, systolic, diastolic.
In the
Setup NBP menu, select Al. from and choose from:
If
Mean is not selected as alarm source (Sys., Dia., or Sys & Dia selected), but the monitor can only derive
a mean value, mean alarms will nevertheless be announced using the most recent mean alarm limits.
Check that the mean alarm limits are appropriate for the patient, even when not using mean as the
alarm source. When no value can be derived an
NBP Measure Failed INOP will be displayed.
Menu option Pressure value monitored
Sys. systolic
Dia. diastolic
Mean mean
Sys & Dia systolic and diastolic in parallel
Dia & Mean diastolic and mean in parallel
Sys & Mean systolic and mean in parallel
Sys&Dia&Mean all three pressures in parallel