24 Specifications
293
Monitor Performance Specifications
Performance Specifications
Power Specifications Power consumption <12 W average
<30 W while battery is loading
Operating Voltage 36 to 60 V DC floating
Battery Specifications Operating Time (with new, fully charged
battery at 25°C)
Basic monitoring configuration: 2.5 hours
(
Brightness set to Optimum, ECG/Resp, SpO
2
measurements
in use, NBP measurement every 15 minutes)
Charge Time When monitor is off: 2 hours
When monitor is in use, and connected to the external power
supply (M8023A), without MMS extensions: 12 hours approx.
Indicators Alarms Off red or yellow LED with crossed out alarms symbol
Alarms red/yellow/light blue (cyan) LED
On/Standby / Error green / red LED
AC Power green LED
Battery yellow (charging)/red blinking (empty) LED
External Power green LED
Sounds Audible feedback for user input
Prompt tone
QRS tone, or SpO
2
modulation tone
4 different alarm sounds
Display Wave Speeds 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
with ±5% accuracy (guaranteed only for integrated displays)
Trends Resolution 12 or 16 numerics @ 12 sec, 1 minute, 5 minute resolution.
Information Multiple choices of number of numerics, resolution and duration
depending on trend option and application area.
For example:
For neonatal, you can choose between 12 and 16 numerics.
Alarm signal System alarm delay
The system alarm delay is the processing
time the system needs for any alarm to be
indicated on the monitor, after the
measurement has triggered the alarm.
less than 4 seconds
Delay for alarm availability on the network
This is the time needed after alarm
indication on the monitor until the alarm
signal is available on the network, to the
IntelliVue Information Center or for
transmission to other systems.
less than 5 seconds
Pause duration 1,2,3 minutes or infinite, depending on configuration
Extended alarm pause 5 or 10 minutes
Sound pressure range min. 0 dB(A)
max. 45-85 dB(A)
Review Alarms Information all alarms / inops, main alarms on/off, alarm silence and time of
occurrence
Capacity 300 items