24 Specifications
314
In this table, U
T
is the a.c. mains voltage prior to application of the test level.
Recommended Separation Distance
WARNING
This equipment generates, uses and radiates radio-frequency energy, and if it is not installed and
used in accordance with its accompanying documentation, may cause interference to radio
communications.
The monitor, equipped with a wireless network interface, intentionally receives RF electromagnetic
energy for the purpose of its operation. Therefore, other equipment may cause interference, even
if that other equipment complies with CISPR emission requirements.
In the following table, P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in meters (m). The
values given in brackets are for respiration.
Portable and mobile RF communications equipment should be used no closer to any part of the
monitor, including cables, than the recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter.
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be
less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with this symbol:
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be
a t levels characteristic of a typical location
in a typical commercial and/or hospital
environment
Immunity test
IEC 60601-1-2
test level
Compliance level Electromagnetic environment guidance
Immunity test
IEC 60601-1-2
test level
Compliance level Electromagnetic environment guidance
Conducted RF
IEC 61000-4-6
3Vrms
150 kHz to 80 MHz
3Vrms
(1 Vrms for respiration)
Recommended separation distance:
d = 1.2P
for respiration:
d = 3.5P
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to 2.5 GHz
3V/m
(1 V/m for respiration)
For short range radio, see
note
1
.
Recommended separation distance:
80 MHz to 800 MHz: d = 1.2P
80 MHz to 800 MHz for respiration: d = 3.5P
800 MHz to 2,5 GHz: d = 2.3P
800 MHz to 2,5 GHz for respiration: d = 7.0P
2.0 to 2,3 GHz for short range radio: d = 7.0P