4 Patient Alarms and INOPs
68
Technical Alarm Messages (INOPs)
If an INOP interrupts monitoring and alarm detection, the measurement numeric will be replaced
by
-?-. If an INOP may lead to unreliable measurement values, a ? appears next to the numeric.
The measurement labels and abbreviations for pressure, temperature and SpO
2
INOP messages are
explained in the individual measurement chapters.
Monitor INOPs
INOP Message, Indication What to do
Bad MSL
INOP tone
1) An MMS with an incompatible software revision is connected to the monitor. This combination
does not allow monitoring, OR
2) You cannot use this combination of monitor, MMS and cable. Switch off the monitor and contact
your service personnel.
Central: Tele Only
INOP tone
System connectivity via telemetry device is limited (No alarms, only local numerics) when in
companion mode and host monitor does not have system connectivity. Only telemetry device
parameters can be displayed at central station.
Check Alarm Lamps
INOP tone
Perform a visual check of the alarm lamp to establish whether there is a problem. Contact your service
personnel to check the internal connections to the alarm lamps.
Check DrugSettings
INOP tone
There was a problem loading the drug settings. Check that the settings are complete and correct.
!!Check ECG Source
Yellow tone
The telemetry device and the monitor both have valid ECG signals. Unpair the telemetry device and
the monitor if they are no longer used for the same patient.
!! Check Equipment
INOP tone/Yellow tone
There is an equipment status dispute relating to one or more of the devices assigned to this patient. See
the Equipment window for details.
Check Flex Texts
INOP tone
Check the names of the monitor menus, for example the labels for screens, profiles, event or trend
group names, before you resume monitoring. If they are unexpected, there may be a problem with the
monitor software. Contact your service personnel.
Check Keyboard
INOP tone
Perform a visual and functional check of the keyboard. Contact your service personnel.
Check Main Board 2
INOP tone
There is a problem with the second main board in the monitor. Contact your service personnel.
Check Monitor Func
INOP tone
Potential problem with alarm lamps, display or interfaces detected. Contact your service personnel.
Check Monitor Temp
INOP tone
The temperature inside the monitor is too high. Check that the monitor ventilation is not obstructed.
If the situation continues, contact your service personnel.
Check Mouse
INOP tone
Perform a visual and functional check of the mouse input device. Contact your service personnel.
Check MSL Voltage
INOP tone
There is a problem with the voltage of the Measurement Link (MSL). Contact your service personnel.
Check Network Conf
INOP tone
The monitor is receiving network topology information from more than one source, e.g. the Database
Server and an Application Server. Contact your service personnel.
!! Check Pairing
Yellow tone
There is a problem with device pairing. Check that the monitor and telemetry device are correctly
paired.
!!Check Patient ID
Yellow tone
There is a mismatch between patient data in two connected devices. Resolve the mismatch to allow
settings and data synchronization.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips MP2 and is the answer not in the manual?

Philips MP2 Specifications

General IconGeneral
BrandPhilips
ModelMP2
CategoryMedical Equipment
LanguageEnglish

Related product manuals