25 Default Settings Appendix
322
SpO2 Default Settings
NBP Default Settings
SpO
2
Settings
Factory Defaults
Adult Pedi Neo
Mode Continuous Continuous Continuous
Repeat Time
15 min 15 min 15 min
Alarms On On On
QRS Volume
111
Tone Modulation Yes Yes Yes
Tone Mod. Type Enhanced Enhanced Enhanced
Perfusion OnOnOn
Average
10 sec 10 sec 10 sec
NBP Alarm Suppr. OnOnOn
Extd. Auto OnOff Disabled Disabled Disabled
Color Cyan
(light blue) Cyan (light blue) Cyan (light blue)
Average in Mon. No No No
Signal Quality OnOnOn
SpO
2
Alarm Default Settings
Setting Adult Pedi Neo
Desat Limit
80 80 80
Low Limit 90 90 85
High Limit 100 100 95
Desat Delay 20 sec 20 sec 20 sec
High Alarm Delay 10 sec 10 sec 10 sec
Low Alarm Delay 10 sec 10 sec 10 sec
SmartAlarmDelay (SAD) Off Off Off
High Alarm Delay
(SAD) Short Short Short
Low Alarm Delay
(SAD) Short Short Short
Label SpO₂ SpO₂ SpO₂
Pulse settings
Pulse (SpO₂) OnOnOn
For all other Pulse default settings, refer to the Pulse Default Settings table.
NBP Settings
Factory Defaults
Adult Pedi Neo
Mode Auto Auto Manual
Alarms from Sys. Sys. Sys.
High Limit
160/90 (110) 120/70 (90) 90/60 (70)