Avoiding Electromagnetic Interference (Resp)
The respiration (Resp) measurement is a very sensitive measurement that measures a very small signal.
Technological limitations do not allow higher immunity levels than 1 V/m for radiated RF
electromagnetic fields and 1 Vrms for conducted disturbances induced by RF fields. Electromagnetic
fields with field strengths above 1 V/m and conducted disturbances above 1 Vrms may cause
erroneous measurements. Therefore Philips recommends that you avoid using electrically radiating
equipment in the close proximity of this measurement.
WARNING
The monitor should not be used next to or stacked with other equipment. If you must stack the
monitor, you must check that normal operation is possible in the necessary configuration before you
start monitoring patients.
Electromagnetic Immunity
The monitor is suitable for use in the specified electromagnetic environment. The user must ensure
that it is used in the appropriate environment as described below.
RF emissions CISPR 11 Class B The monitor, with the following measurements and interfaces:
ECG/Respiration, NBP, SpO
2
1
, Pressure, Temperature, CO
2
(only Mainstream
Sensor M2501A), LAN
is suitable for use in all establishments including those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Emissions test Compliance Avoiding Electromagnetic Interference
Immunity test
IEC 60601-1-2
test level
Compliance level Electromagnetic environment guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a
typical commercial and/or hospital
environment
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be that of a
typical commercial and/or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycles
<5% U
T
(>95% dip in U
T
) for
0.5 cycles
Mains power quality should be that of a
typical commercial and/or hospital
environment. If the user of the monitor
requires continued operation during power
mains interruptions, it is recommended
that the monitor is equipped with an
internal battery or is powered from an
uninterruptible power supply.
40% U
T
(60% dip in U
T
) for
5cycles
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for
5sec
<5% U
T
(>95% dip in U
T
) for
5sec