14 Monitoring Carbon Dioxide
211
Setting up CO2 Corrections
Temperature, water vapor in the patient's breath, barometric pressure, and the proportions of O
2,
N
2
O
and Helium in the mixture all influence CO
2
absorption. If values seem inaccurately high or low, check
that the monitor is using the appropriate corrections.
Suppressing Sampling (not Mainstream CO2)
To temporarily stop sampling, in the Setup CO menu, select Pump 15min Off or use the CO Pump Off
SmartKey if available.
Selecting
Pump 15min Off again before fifteen minutes have passed resets the timer to fifteen minutes.
To re-start the pump, in the
Setup CO menu, select Pump On.
Changing CO2 Alarms
This refers to CO
2
specific alarms. See the Alarms section for general alarm information.
To change the etCO
2
alarm limits, in the Setup CO menu:
1 Touch the high limit on the alarm limits menu key.
2 Choose the high alarm limit from the pop-up list that opens.
3 Touch the low limit on the alarm limits menu key.
4 Choose the low alarm limit from the pop-up list that opens.
Correction
Altitude
(M3014A only)
Altitude is set during installation. The monitor automatically applies an
appropriate correction.
O
2
(M3014A only)
In the
Setup CO menu, select Oxy. Corr and select a value between 20% and
100%, the default value is 20%.
Humidity At installation, the monitor is configured to automatically apply either Body
Temperature Pressure Saturated (BTPS) or Ambient Temperature Pressure
Dry (ATPD). To see which, go to the
Setup CO menu, and scroll down to
look at
Hum. Corr or Humidity Corr..
N
2
O
(M3015A/B
microstream only)
In the
Setup CO menu, select NO Corr. and turn on or off. If N
2
O is present
in the ventilation gas mixture, you must turn this on.
If the N
2
O correction is not available in the Setup CO menu, the CO
2
measurement in your MMS Extension does not require N
2
O correction or it
is setup with
Gas Corr. (see below).
Gas
(M3014A only)
In the
Setup CO menu, select Gas Corr. and select Helium, NO or turn off. If
Helium or N
2
O is present in the ventilation gas mixture, you must make the
appropriate selection.
If the Gas correction is not available in the
Setup CO menu, the CO
2
measurement in your MMS Extension does not require N
2
O or Helium
correction or the N
2
O correction is setup with NO Corr. (see above).
Agent
(M3014A only)
In the
Setup CO menu, select Agt. Corr and select the concentration of the
anesthetic agent (between 0.0% and 20.0%). If an anesthetic agent is present
in the ventilation gas mixture, you must select the appropriate concentration.