24 Specifications
302
Philips FAST SpO2 Performance Specifications
Philips FAST SpO
2
Performance Specifications
SpO2
The specified accuracy
is the root-mean-
square (RMS)
difference between the
measured values and
the reference values
Range 0 to 100%
Accuracy Philips Reusable Sensors:
M1191A, M1191AL, M1191B, M1191BL, M1192A, = 2% (70% to 100%)
M1193A, M1194A, M1195A, M1196A, M1196S = 3% (70% to 100%)
M1191T, M1192T, M1193T (Adult), M1196T = 3% (70% to 100%)
M1193T (Neonate) = 4% (70% to 100%)
Philips Disposable Sensors with M1943A(L):
M1132A, M1133A (adult/infant), M1134A (adult/infant) = 2%
M1901B, M1902B, M1903B, M1904B, M1131A, M1133A (neonate), M1134A (neonate) =
3% (70% to 100%)
NellcorPB
®
Sensors with M1943A(L):
MAX-A, MAX-AL, MAX-P, MAX-I, MAX-N, D-25, D-20, I-20, N-25, OxiCliq A, P, I, N
= 3% (70% to 100%)
Masimo Reusable Sensors
®
with LNOP MP12 or LNC MP10:
LNOP DC-I, LNOP DC-IP, LNOP YI (adult/pedi/infant), LNCS DC-1, LNCS DC-IP,
LNCS YI (adult/pedi/infant) = 2% (70% to 100%)
LNOP YI (neonate), LNCS YI (neonate) = 3% (70% to 100%)
LNOP TC-I, LNCS TC-I: 3.5% (70% to 100%)
Masimo Disposable Sensors
®
with LNOP MP12 or LNC MP10:
LNOP Adt, LNOP Adtx, LNOP Pdt, LNOP Pdtx, LNOP Inf-L, LNOP Neo-L (adult),
LNCS Adtx, LNCS Adtx-3, LNCS Pdtx, LNCS Pdtx-3, LNCS Inf, LNCS Inf-3, LNCS
Neo (adult), LNCS Neo-3 (adult) = 2% (70% to 100%)
LNOP Neo-L (neonate), LNOP NeoPt-L, LNCS Neo (neonate), LNCS Neo-3 (neonate),
LNCS NeoPt, LNCS NeoPt-3 = 3% (70% to 100%)
Resolution 1%
Pulse Range 30 to 300 bpm
Accuracy ±2% or 1 bpm, whichever is greater
Resolution 1 bpm
Numeric Update Rate Typical: 2 seconds, Maximum: 30 seconds. Maximum with NBP INOP suppression on: 60
seconds.
Sensors Wavelength range: 500 to 1000 nm
Emitted Light Energy: ≤15 mW
Information about the wavelength range can be especially useful to clinicians (for instance,
when photodynamic therapy is performed)
Pulse Oximeter Calibration Range 70% to 100%