24 Specifications
304
Nellcor Patents: The monitor with Nellcor OxiMax technology are covered by one or more of the
following US Patents and foreign equivalents:
5,485,847; 5,676,141; 5,743,263; 6,035,223; 6,226,539, 6,411,833; 6,463,310; 6,591,123; 6,708,049;
7,016,715; 7,039,538; 7,120,479; 7,120,480; 7,142,142; 7,162,288; 7,190,985; 7,194,293; 7,209,774;
7,212,847; 7,400,919.
Purchase of the instrument with Nellcor OxiMax technology confers no express or implied license
under any Covidien patent to use that instrument with any sensor not manufactured or licensed by
Covidien.
SpO2 Alarm Specifications
1. Refer to “Monitor Performance Specifications” on page 293 for system alarm delay specification.
SpO
2
Alarm Specifications Range Adjustment
SpO
2
Adult: 50 to 100%
Pedi/Neo: 30 to 100%
1% steps
Desat Adult: 50% to Low alarm limit
Pedi/Neo: 30% to Low alarm limit
1% steps
Pulse 30 to 300 bpm Adult:
1 bpm steps (30 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (30 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Tachycardia Difference to high limit 0 to 50 bpm 5 bpm steps
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to 50 bpm 5 bpm steps
Clamping at 30 to 100 bpm 5 bpm steps
SpO
2
Alarm
Specifications (cont.)
Standard Delay
Smart Alarm Delay (alternative to standard alarm delay)
Short Mode Medium mode Long Mode
SpO
2
high and low limit
alarms
0 to 30 seconds (adjustable in
1 second steps) + system alarm
delay
1
10 to 25 seconds +
system alarm delay
10 to 50 seconds
+system alarm delay
10 to 100 seconds +
system alarm delay
Desat 0 to 30 seconds (adjustable in
1 second steps) + system alarm
delay
not supported
Pulse 10 seconds + system alarm delay
Tachycardia 10 seconds + system alarm delay
Bradycardia 10 seconds + system alarm delay