1 Basic Operation
29
Understanding Screens
Your monitor comes with a set of pre-configured Screens, optimized for common monitoring
scenarios such as "OR adult", or "ICU neonatal". A Screen defines the overall selection, size and
position of waves, numerics and other elements on the monitor screen when you switch on. You can
easily switch between different Screens during monitoring. Screens do NOT affect alarm settings,
patient category and so forth.
Switching to a Different Screen
To switch to a different Screen:
1 After closing any open menus or windows, press the Main Screen key to access the Change Screen
menu.
2 Choose the new Screen from the Change Screen menu.
Changing a Screen’s Content
If you do not want to change the entire Screen content, but only some parts of it, you can substitute
individual waves, numerics, or trends. Be aware that these changes cannot be stored permanently in
Monitoring Mode.
To change the selection of elements on a Screen,
1 Select the element you want to change. For example, touch the wave to enter the wave setup menu,
or touch the numeric to enter the numeric setup menu.
2 From the menu that appears, select Change Wave or Change Numeric, and then select the wave or
numeric you want.
In the
Change Screen menu, the changed Screen is shown linked to the original Screen and marked with
an asterisk.
Up to three modified Screens can be accessed via the
Change Screen menu.
To recall Screens, select the name of the Screen in the
Change Screen menu.
After a patient discharge, the monitor’s default Screen is shown. Modified Screens are still available in
the
Change Screen menu.
If the monitor is switched off and then on again, modified Screens are erased from the monitor’s
memory and cannot be recalled. If a modified Screen was the last active Screen when the monitor was
switched off, it is retained unless the monitor is configured to revert to the default.
Using the XDS Remote Display
Using the IntelliVue XDS Solution it is possible to view an independent monitor screen on an external
display. The XDS Solution consists of a medical grade PC-based hardware platform, XDS Application
software and the XDS connectivity option on the monitor. Depending on the configuration you can
also operate the monitor from the external display. The XDS must be connected to the same Local
Area Network (LAN) as the monitor.
It is also possible to use an existing PC, connected to the same LAN, to host the XDS Application
software.
For more details, including limitations and restrictions, refer to the IntelliVue XDS Application
Instructions for Use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips MP2 and is the answer not in the manual?

Philips MP2 Specifications

General IconGeneral
BrandPhilips
ModelMP2
CategoryMedical Equipment
LanguageEnglish

Related product manuals