EasyManuals Logo

Philips MP2 User Manual

Philips MP2
334 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #325 background imageLoading...
Page #325 background image
25 Default Settings Appendix
325
CO2 Default Settings
Unit mmHg mmHg mmHg
Color Yellow Yellow Yellow
PAP Settings
Factory Defaults
Adult Pedi Neo
ICP, IC1, IC2 Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit
14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Enabled Enabled Enabled
Δ Extreme High
10 10 10
Δ Extreme Low 10 10 10
Mean Low Clamp 20/10 (-5) 15/5 (-5) 15/5 (-5)
Mean High Clamp 0/-5 (0) 0/-5 (0) 0/-5 (0)
Scale 30 30 30
Mean only Yes Yes Yes
Filter
12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr.
60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Magenta Magenta Magenta
CO
2
Settings
Factory Defaults
Adult Pedi Neo
etCO Low
25 25 25
etCO High 60 60 60
imCO High 444
CO Alarms On On On
Unit
mmHg mmHg mmHg
Scale 50 mmHg 50 mmHg 50 mmHg
imCO On On On
NO Corr.
(only M3015A/M3016A) Off Off Off
Oxygen Corr.
(only M3014A) 20% 20% 20%
Gas Corr. (only M3014A) Off Off Off
Humidity Corr. BTPS BTPS BTPS
Max Hold
20 s 20 s 20 s
awRR On On On