25 Default Settings Appendix
324
Mercury Cal Yes Yes Yes
Artifact Suppr.
60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Red Red Red
ABP, ART, Ao, BAP, FAPP,PP1, P2, P3, P4, UAP
Settings
Factory Defaults
Adult Pedi Neo
CVP, RAP, LAP, UVP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Mean Mean Mean
High Limit
14/6 (10) 10/2 (4) 10/2 (4)
Low Limit 6/-4 (0) 2/-4 (0) 2/-4 (0)
Alarms On On On
Extreme Alarms Enabled Enabled Enabled
Δ Extreme High
555
Δ Extreme Low 555
Mean High Clamp 20/10 (15) 15/5 (10) 15/5 (10)
Mean Low Clamp 0/-5 (-5) 0/-5 (-5) 0/-5 (-5)
Scale 30 30 30
Mean only Yes Yes Yes
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr.
60 sec 60 sec 60 sec
Unit mmHg mmHg mmHg
Color Cyan
(light blue) Cyan (light blue) Cyan (light blue)
PAP Settings
Factory Defaults
Adult Pedi Neo
Alarms from Dia. Dia. Dia.
High Limit
34/16 (20) 60/4 (26) 60/4 (26)
Low Limit 10/0 (0) 24/-4 (12) 24/-4 (12)
Alarms On On On
Extreme Alarms Enabled Enabled Enabled
Δ Extreme High
555
Δ Extreme Low 555
Dia. High Clamp 45/20 (25) 65/5 (35) 65/5 (35)
Dia. Low Clamp 5/-5 (-5) 15/-5 (5) 15/-5 (5)
Scale 30 30 30
Mean only No No No
Filter 12 Hz 12 Hz 12 Hz
Mercury Cal Yes Yes Yes
Artifact Suppr.
60 sec 60 sec 60 sec