Invasive Pressure Alarm Specifications
Temp
Complies with EN 12470-4:2000. Specified without transducer.
Invasive Pressure Alarm
Specifications
Range Adjustment Delay
Pressure -40 to 360 mmHg
(-5.0 to 48 kPa)
-40 to 50 mmHg
2mmHg (0.5kPa)
>50 mmHg
5mmHg (1kPa)
max. 12 seconds
Extreme High Difference to high limit 0 to
25 mmHg
5 mmHg steps (0.5 kPa)
Clamping at -40 to 360 mmHg 5 mmHg steps (1.0 kPa)
Extreme Low Difference to low limit 0 to
25 mmHg
5 mmHg steps (0.5 kPa)
Clamping at -40 to 360 mmHg 5 mmHg steps (1.0 kPa)
Pulse 25 to 300 bpm Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Tachycardia Difference to high limit 0 to
50 bpm
5 bpm steps max. 14 seconds
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to
50 bpm
5 bpm steps max. 14 seconds
Clamping at 25 to 100 bpm 5 bpm steps
Temp Performance Specifications
Temp Range -1 to 45°C (30 to 113°F)
Resolution 0.1°C (0.2°F)
Accuracy ±0.1°C (±0.2°F)
Average Time Constant Less than 10 seconds
Temp Alarm Specifications Range Adjustment
Temp High/Low Alarms -1 to 45°C (30 to 113°F) -1 to 30°C (30 to 86°F), 0.5°C (1.0°F) steps
30 to 45°C (86 to 113°F), 0.1°C (0.2°F) steps