Pulse Default Settings
Pulse alarms use the settings of the currently selected Pulse alarm source.
Respiration Default Settings
Pulse Settings
Factory Defaults
Adult Pedi Neo
Alarm Source Auto Auto Auto
Pulse
(source label) On On On
System Pulse SpO₂ SpO₂ SpO₂
Alarms Off Enabled Enabled Enabled
AlarmSource Sel. Enabled Enabled Enabled
Pulse Alarm Settings
Factory Defaults H10/H20/H40
Adult Pedi Neo
Pulse (SpO₂)OnOnOn
Pulse Alarms Off Off Off
High Limit
120 bpm 160 bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Δ ExtrBrady 20 bpm 20 bpm 20 bpm
Brady Clamp 40 bpm 40 bpm 50 bpm
Δ ExtrTachy 20 bpm 20 bpm 20 bpm
Tachy Clamp 200 bpm 220 bpm 240 bpm
Respiration Settings
Factory Defaults
Adult Pedi Neo
High Limit
30 rpm 30 rpm 100 rpm
Low Limit 8 rpm 8 rpm 30 rpm
Apnea Time 20 sec 20 sec 20 sec
Alarms On On On
Resp On On On
Detection Auto
(Trigger Mode)
Auto
(Trigger Mode)
Auto
(Trigger Mode)
Color Yellow Yellow Yellow