25Default Settings Appendix
This appendix documents the most important default settings of your monitor as it is delivered from
the factory. For a comprehensive list and explanation of default settings, see the Configuration Guide
supplied with your monitor. The monitor's default settings can be permanently changed in
Configuration Mode.
NOTE
If your monitor has been ordered pre-configured to your requirements, the settings at delivery will be
different from those listed here.
Alarm and Measurement Default Settings
The default wave speed for all measurements except Respiration is 25 mm/sec. For Respiration the
default wave speed is 6.25 mm/sec.
In the following tables, settings may be entered only once per table row if they are the same for all
patient categories.
Alarm Default Settings
Alarm Settings Factory Default H10/H20/H40
H30/H31 (deviations from H10/H20/
H40)
Alarm Volume 5
Alarms Off 2 min
Pause Al. 5min Enabled
Pause Al. 10min Enabled
Auto Alarms Off Off
AlarmOffReminder Off
Visual Latching Red&Yellow Red Only
Audible Latching Red&Yellow Off
Alarm Reminder On
Reminder Time 3 min
Alarm Sounds Traditional
RedAlarmInterval
10 sec
Yel.Al. Interval 20 sec