6 ECG, Arrhythmia, ST and QT Monitoring
113
Standard 3-Lead Placement
Standard 5-Lead Placement
V4 C4 Brown/Blue White/Brown
V5 C5 Brown/Orange White/Black
V6 C6 Brown/Violet White/Violet
Electrode labels Electrode colors
AAMI EASI IEC AAMI IEC
1 RA placement: directly below the clavicle and
near the right shoulder
2 LA placement: directly below the clavicle and
near the left shoulder
3 LL placement: on the left lower abdomen
1 RA placement: directly below the clavicle and
near the right shoulder
2 LA placement: directly below the clavicle and
near the left shoulder
3 RL placement: on the right lower abdomen
4 LL placement: on the left lower abdomen
5 V placement: on the chest, the position
depends on your required lead selection. If
configured, the label of the chest lead will be
displayed with its exact position (V1 to V6R)