24 Specifications
305
NBP
Complies with IEC 60601-2-30:1999/EN60601-2-30:2000.
SpO
2
Smart Alarm Delay - Detailed Specifications
Deviation from violated
alarm limit
Resulting alarm delay according to selected mode
Short Medium Long
1% 25 seconds
(maximum delay)
50 seconds
(maximum delay)
100 seconds
(maximum delay)
2% 12 seconds 25 seconds 50 seconds
3% 10 seconds 16 seconds 33 seconds
4% 10 seconds 12 seconds 25 seconds
5% 10 seconds 10 seconds 20 seconds
6% 10 seconds 10 seconds 16 seconds
7% 10 seconds 10 seconds 14 seconds
8% 10 seconds 10 seconds 12 seconds
9% 10 seconds 10 seconds 11 seconds
>9% 10 seconds 10 seconds 10 seconds
NBP Performance Specifications
Measurement Ranges Systolic Adult: 30 to 270 mmHg (4 to 36 kPa)
Pedi: 30 to 180 mmHg (4 to 24 kPa)
Neo: 30 to 130 mmHg (4 to 17 kPa)
Diastolic Adult: 10 to 245 mmHg (1.5 to 32 kPa)
Pedi: 10 to 150 mmHg (1.5 to 20 kPa)
Neo: 10 to 100 mmHg (1.5 to 13 kPa)
Mean Adult: 20 to 255 mmHg (2.5 to 34 kPa)
Pedi: 20 to 160 mmHg (2.5 to 21 kPa)
Neo: 20 to 120 mmHg (2.5 to 16 kPa)
Pulse Rate Adult: 40 to 300 bpm
Pedi: 40 to 300 bpm
Neo: 40 to 300 bpm
Accuracy Max. Std. Deviation: 8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate Measurement Accuracy 40 to 100 bpm: ±5 bpm
101 to 200 bpm: ±5% of reading
201 to 300 bpm: ±10% of reading
(average over NBP measurement cycle)
Pulse Rate Range 40 to 300 bpm