4 Patient Alarms and INOPs
67
** <Temp Label> High Temp The temperature has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** <Temp Label> Low Temp The temperature has fallen below the low alarm
limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
* T-Mon Alarm
** T-Mon Alarm
*** T-Mon Alarm
an assigned
transport monitor
A transport monitor used for the patient is in
an alarm condition. Check the detailed alarm
information at the Information Center.
yellow or red alarm lamp and alarm
tone
** TOFcnt High NMT TOFcnt value has exceeded the high alarm
limit
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
* Vent Bigeminy ECG/Arrhythmia A dominant rhythm of N, V, N, V (N =
supraventricular beat, V = ventricular beat)
numeric flashes, yellow alarm lamp,
short yellow audible alarm
*** Vent Fib/Tach ECG Fibrillatory waveform (sinusoidal wave between
2 - 10 Hz) for 4 consecutive seconds.
numeric flashes, red alarm lamp,
alarm tone
* Vent Rhythm
** Vent Rhythm
ECG/Arrhythmia A dominant rhythm of adjacent Vs > the vent
rhythm limit and ventricular heart rate < the V-
Tach HR limit.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
* Vent Trigeminy ECG/Arrhythmia A dominant rhythm of N, N, V, N, N, V (N =
supraventricular beat, V = ventricular beat).
numeric flashes, yellow alarm lamp,
short yellow audible alarm
*** VTach ECG, Arrhythmia A run of PVCs the V-Tach run limit and
heart rate > the V-Tach HR limit.
numeric flashes, red alarm lamp,
alarm tone
** VueLink Alarm
*** VueLink Alarm
at Information Center
VueLink A yellow (**) or red (***) patient alarm is
present on the VueLink module. Check the
monitor display for more detailed alarm
information.
(on monitor) yellow or red alarm
lamp; the alarm text is defined by
the VueLink device driver
***xBrady xxx<yyy Press, SpO
2,
ECG This is the alternative alarm message for
*** Extreme Brady, when the alarm text
setting is
Enhanced.
Heart rate < the extreme bradycardia alarm
limit. xxx denotes the lowest measured value;
yyy is the extreme bradycardia limit.
numeric flashes and alarm limit is
highlighted, red alarm lamp, alarm
tone
***xTachy xxx>yyy Press, SpO
2
, ECG This is the alternative alarm message for ***
Extreme Tachy
when the alarm text setting is
Enhanced.
Heart rate >the extreme tachycardia alarm
limit. xxx denotes the highest measured value;
yyy is the tachycardia limit.
numeric flashes, alarm limit is
highlighted, red alarm lamp, alarm
tone