4 Patient Alarms and INOPs
69
Check Screen Res.
INOP tone
The Screen you have selected uses a resolution which is not supported by the display. The monitor will
show a generic Screen instead until you select a different Screen.
Contact your service personnel if you want the Screen deleted from the Profile(s) to avoid this in
future.
Check Settings
INOP tone
If this INOP appears, check the monitor and patient settings before you resume monitoring. If the
settings are unexpected, there may be a problem with the monitor software. Contact your service
personnel.
Check SpeedPoint
INOP tone
Perform a visual and functional check of the SpeedPoint input device. Contact your service personnel.
Check Touch Input
INOP tone
Perform a visual and functional check of the touch input device. Contact your service personnel.
Check Waves
INOP tone
The options purchased with this monitor may not support the number of waves required to show the
selected Screen, so some waves or high resolution trends are missing from the Screen. Select a
different Screen with fewer waves.
Contact your service personnel if you want the Screen deleted from the Profile(s) to avoid this in
future.
Chk ECG Sync Cable
INOP tone
The ECG Sync is detecting an invalid signal, or the ECG Sync cable is disconnected.
Chk IndepDsp Cable The monitor cannot communicate with the D80 Intelligent Display. Check the MSL coupling cable.
The end with the gray connector must be connected to the Intelligent Display.
Chk <Measurement
Label>Settings
Synchronization of the settings for the measurement cited, has failed. Check that settings are
appropriate for your patient.
Chk MSL Connection
INOP tone
Check that the MSL cable is properly connected. If this is the case, try using another MSL cable, to
check if your cable is defective. If this does not help, the device connected via the MSL cable may be
defective, contact your service personnel.
Internal.Comm.Malf
INOP tone
There is a problem with I2C Bus communication in the monitor. Contact your service personnel.
MCC Reversed
INOP tone
The MSL coupling cable is reversed. Connect the end with the gray connector to the Intelligent
Display.
MCC Unsupported
INOP tone
An MSL coupling cable has been connected to a device which does not support MSL coupling.
Meas. Deactivated An X2 has been connected to a host monitor (companion mode) and all derived measurements have
been deactivated and/or measurements with a label conflict. The measurements can only be
reactivated by disconnecting the measurement device from the host monitor.
!!More Bed Alarms
!!!More Bed Alarms
at Information Center
The monitor is associated with a telemetry device and is sending data to the Information Center via
the telemetry device. There are currently more alarms at the bedside than can be transmitted to the
Information Center.
More INOPs
!! More INOPs
!!! More INOPs
There is more than one technical alarm of the corresponding severity active on the devices that
monitor this patient. For details check the information displayed at the Information Center.
MSL Power High The power consumption of the devices connected to the Measurement Link (MSL) cable is too high.
If this situation continues, the MSL will be switched off. Contact your service personnel.
MSL Power Off
INOP tone
The power consumption of the devices connected to the Measurement Link (MSL) cable was too high
for too long and the MSL has been switched off. Contact your service personnel.
MSL Power Overload
INOP tone
The power consumption of the devices connected to the Measurement Link (MSL) cable is much too
high or there has been a short circuit. The MSL has been switched off. Contact your service personnel.
INOP Message, Indication What to do