4 Patient Alarms and INOPs
70
No Central Monit.
INOP tone
There is a problem with the communication to the network. Central monitoring is currently not
possible (no patient alarms or information). Check the connection. In case the connection is via a
telemetry device, the current telemetry use model does not support central monitoring. Contact your
service personnel.
No ECG at Central The ECG measured with the monitor ECG is not being sent to the Information Center via the
telemetry device.
OtherMon INOP
!! OtherMon INOP
!!! OtherMon INOP
Another monitor assigned to the patient (e.g. a transport monitor) is in an INOP condition. Check the
detailed INOP information at the Information Center and resolve the INOP condition(s) at the other
monitor.
Rem.AlarmDev.Malf.
INOP tone
There is a problem with the connection to the remote alarm device. Contact your service personnel to
check the remote alarm device and its connections.
Settings Malfunct
INOP tone
The monitor cannot use the predefined settings for monitoring. Contact your service personnel.
Speaker Malfunct
INOP tone
Contact your service personnel to check the speaker and the connection to the speaker.
SRR Incompatible
INOP tone
The short range radio component has compatibility problems. Contact your service personnel.
SRR Interference
INOP tone
The short range radio connection has interference from another device. Try using another channel.
SRR Invalid Chan
INOP tone
The channel configuration of the Short Range Radio is invalid. Check channel and channel mask
configuration.
SRR Malfunction
INOP tone
Malfunction in the short range radio device. If the INOP persists contact your service personnel.
TAAP Disabled
INOP tone
The currently selected telemetry configuration on the monitor does not allow connection of telemetry
devices to the monitor.
Tele Config Unsupp
INOP tone
Telemetry Workflow configuration not supported. Check revision and configuration of monitor and
central.
Tele Equip Malf
INOP tone
The telemetry device has a malfunction. Disconnect and reconnect the telemetry device. If the INOP
reappears, replace the telemetry device.
Tele Incompatible SRR-enabled telemetry device is not supported by this central software revision. Please check
configuration.
Tele INOP
!! Tele INOP
!!! Tele INOP
A telemetry device assigned to the patient is in an INOP condition. Check the detailed INOP
information at the Information Center and resolve the INOP condition(s) at the telemetry device.
Tele Unsupported
INOP tone
This telemetry device is not supported for direct connection to the monitor.
<Timer Label> Elapsed
INOP tone
The time has expired for the timer indicated in the INOP text. Clearing the timer clears the INOP.
T-Mon INOP
!! T-Mon INOP
!!! T-Mon INOP
INOP tone
A transport monitor assigned to the patient is in an INOP condition. Check the detailed INOP
information at the Information Center and resolve the INOP condition(s) at the transport monitor.
TransportEquipment The equipment is ready for transport. See the Equipment window for details.
Unsupported LAN
INOP tone
There is a problem with the communication to the network and central monitoring is currently not
possible. Check the connection. If the INOP persists, switch off the monitor and contact your service
personnel.
User I/F Malfunct
INOP tone
Perform a visual and functional check of all the monitor input devices. Contact your service personnel.
INOP Message, Indication What to do