4 Patient Alarms and INOPs
65
*** <Press Label> High PRESS The measured pressure value is above the
extreme high alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes, high limit is
highlighted, red alarm lamp, alarm
tone
** <Press Label> High PRESS The measured pressure value is above the high
alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes, high limit is
highlighted, yellow alarm lamp,
alarm tone
*** <Press Label> Low PRESS The measured pressure value is below the
extreme low alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes and low limit is
highlighted, red alarm lamp, alarm
tone
** <Press Label> Low PRESS The measured pressure value is below the low
alarm limit.
s, d, or m after the label indicates whether the
systolic, diastolic or mean pressure has crossed
the limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
** <pTemp Label> High pTemp The temperature value has exceeded the high
alarm limit.
numeric flashes and high limit is
high-lighted, yellow alarm lamp,
alarm tone
** <pTemp Label> Low pTemp The temperature value has dropped below the
low alarm limit.
numeric flashes and low limit is
high-lighted, yellow alarm lamp,
alarm tone
** Pulse High PRESS, SpO
2,
cmResp
The pulse rate has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** Pulse Low PRESS, SpO
2
,
cmResp
The pulse rate has dropped below the low
alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
* PVCs/min High ECG/Arrhythmia PVCs detected within a minute > the alarm
limit
numeric flashes, yellow alarm lamp,
short yellow audible alarm
** QTc High ECG/QT QTc value has exceeded the QTc high limit for
more than 5 minutes
numeric flashes, yellow alarm lamp,
alarm tone
** ΔQTc High ECG/QT ΔQTc value has exceeded the ΔQTc high limit
for more than 5 minutes
numeric flashes, yellow alarm lamp,
alarm tone
* R-on-T PVCs ECG/Arrhythmia For HR < 100, a PVC with R-R interval < 1/3
the average interval followed by a
compensatory pause of 1.25*(the average R-R
interval), or two such Vs without compensatory
pause occurring within 5 minutes of each other.
(When HR > 100, 1/3 R-R interval is too short
for beat detection)
numeric flashes, yellow alarm lamp,
short yellow audible alarm
** RR High RESP The respiration rate has exceeded the high
alarm limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** RR Low RESP The respiration rate has dropped below the low
alarm limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication