EasyManua.ls Logo

R&S FSV3000 Series - Page 1026

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1026User Manual 1178.8520.02 ─ 01
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK:FUNC:NDBD ON
Switches on the n dB down function.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK:FUNC:NDBD:RES?
Outputs the measured value.
Usage: Query only
Manual operation: See " n dB down Marker State " on page 452
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:STATe <State>
This command turns the n dB Down marker function on and off.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:FUNC:NDBD:STAT ON
Turns the n dB Down marker on.
Manual operation: See " n dB down Marker State " on page 452
CALCulate<n>:MARKer<m>:FUNCtion:NDBDown:TIME?
This command queries the position of the n dB down markers on the x-axis when mea-
suring in the time domain.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 746.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Return values:
<TimeX1> absolute position in time of the n dB marker to the left of the ref-
erence marker in seconds
<TimeX2> absolute position in time of the n dB marker to the right of the
reference marker in seconds
Analyzing Measurements (Basics)