Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
234User Manual 1178.8520.02 ─ 01
This setting is required to calculate the SEM limits according to standard 3GPP 37.141
V12.2.0.
Remote command:
[SENSe:]ESPectrum<sb>:MSR:GSM:CARRier on page 818
Apply to SEM
Configures the SEM sweep list according to the specified MSR settings.
Remote command:
[SENSe:]ESPectrum<sb>:MSR:APPLy on page 815
7.6.5.6 Standard Files
Access: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "Stan-
dard Files"
You can save the current measurement settings as a user-defined standard (XML file),
or load stored measurement settings. Furthermore, you can delete an existing settings
file.
For details, see Chapter 7.6.6.1, "How to Manage SEM Settings Files", on page 241.
Standard files for sub blocks (Multi-SEM measurements)
If more than one sub blocks are defined, the "Standard Files" tab and softkey are not
available. To load a standard file for an individual sub block, use the Standard / MSR
Settings setting in the "Sub Blocks" tab.
Spectrum Emission Mask (SEM) Measurement