Troubleshooting
R&S
®
FSVA3000/ R&S
®
FSV3000
1215User Manual 1178.8520.02 ─ 01
Device-specific error messages use positive error codes. These messages are
described below.
Table 16-3: Device-specific error messages
Error code Error text in the case of queue poll
Error explanation
1006 Failed to connect to server (code. 1006)
LXI web browser access to the instrument has been deactivated.
1052 Frontend LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails in the RF
front-end.
1060 Trigger-Block Gate Delay Error- gate length < Gate Delay
This message is displayed when the gate signal length is not sufficient for the pull-in delay
with a predefined gate delay.
1064 Tracking LO is Unlocked
This message is displayed when the phase regulation of the local oscillator fails on the
external generator module.
2028 Hardcopy not possible during measurement sequence
This message is displayed when a printout is started during scan sequences that cannot
be interrupted. Such sequences are for example:
Recording the system error correction data (alignment)
Instrument self-test
In such cases synchronization to the end of the scan sequence should be performed prior
to starting the printout.
2033 Printer Not Available
This message is displayed when the selected printer is not included in the list of available
output devices. A possible cause is that the required printer driver is missing or incorrectly
installed.
2034 CPU Temperature is too high
This message is displayed when the temperature of the processor exceeds 70 °C.
Table 16-4: Power Sensor errors
Status bar message Description
Zeroing could not be per-
formed
Zeroing could not be performed because the RF power applied is too
high.
Power sensor zero failed
16.3 Troubleshooting Remote Operation
If problems arise during measurement in remote operation, try the following methods to
solve them.
Troubleshooting Remote Operation

Table of Contents