Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
705User Manual 1178.8520.02 ─ 01
"80 MHz"
The 80 MHz wideband is used.
The option is indicated as "B7J, 122".
Remote command:
SYSTem:PSA:WIDeband on page 1159
FSU/FSQ Preamplifier
This setting defines which option is returned when the *OPT? query is executed,
depending on the used preamplifier.
It is only available for FSU/FSQ emulation, and only if a preamplifier is used by the
R&S FSV/A (-B24/-B25 option).
Remote command:
SYSTem:PREamp on page 1159
HP Additional
Allows the use of HP commands in addition to SCPI commands for R&S
FSP/FSQ/FSU emulation (see Language).
Remote command:
SYSTem:HPADditional on page 1158
Revision String
Defines the response to the REV? query for the revision number.
(HP emulation only, see "Language" on page 703).
Max. 36 characters are allowed.
Remote command:
SYSTem:REVision[:STRing] on page 1160
Resetting the Factory Revision
Resets the response to the REV? query for the revision number to the factory default
(HP emulation only, see "Language" on page 703).
Remote command:
SYSTem:REVision:FACTory on page 1126
13.5.4 LXI Settings
Access: [SETUP] > "Network + Remote" > "LXI" tab
On the R&S FSV/A the LXI Class C functionality is already installed and enabled; thus,
the instrument can be accessed via any web browser (e.g. the Microsoft Internet
Explorer) to perform the following tasks:
●
modifying network configurations
●
modifying device configurations
●
monitoring connections from the device to other devices
The "LXI" tab of the "Network + Remote" dialog box provides basic LXI functions and
information for the R&S FSV/A.
Network and Remote Control Settings