Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
875User Manual 1178.8520.02 ─ 01
-----Initiliazing the measurement and querying results simultaneously-----
MPOW? 935.2MHZ,1MHZ,434us,VID,50,5us,MEAN,20
//Configures, initiates and queries the results of the measurement.
//Result example:
-105.225059509,-105.656074524,-105.423065186,-104.374649048,-103.059822083,-101.29511261,
-99.96534729,-99.7452468872,-99.6610794067,-100.327224731,-100.96686554,-101.450386047,
-102.150642395,-103.240142822,-105.95476532,-110.583129883,-115.7760849,-126.279388428,
-124.620399475,-116.97366333
14.6 Configuring the Result Display
The commands required to configure the screen display in a remote environment are
described here.
The tasks for manual operation are described in Chapter 9.1, "Result Display Configu-
ration", on page 412.
General Window Commands................................................................................ 875
Working with Windows in the Display................................................................... 876
Examples: Configuring the Result Display............................................................882
14.6.1 General Window Commands
The following commands are required to configure general window layout, independent
of the application.
Note that the suffix <n> always refers to the window in the currently selected channel
(see INSTrument[:SELect] on page 739).
DISPlay:FORMat........................................................................................................... 875
DISPlay[:WINDow<n>]:SIZE........................................................................................... 876
DISPlay:FORMat <Format>
This command determines which tab is displayed.
Parameters:
<Format> SPLit
Displays the MultiView tab with an overview of all active chan-
nels
(See Chapter 6.2, "R&S MultiView", on page 112).
SINGle
Displays the measurement channel that was previously focused.
*RST: SING
Example:
DISP:FORM SPL