Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
258User Manual 1178.8520.02 ─ 01
Reference Level
Sets the reference level for the range.
For details on the reference level, see Chapter 8.4.1.1, "Reference Level",
on page 366.
Remote command:
[SENSe:]LIST:RANGe<ri>:RLEVel on page 831
RF Attenuation Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see Chapter 8.4.1.2, "RF Attenuation", on page 368.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation:AUTO on page 829
RF Attenuation
Sets the attenuation value for that range.
For details on attenuation, see Chapter 8.4.1.2, "RF Attenuation", on page 368.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:ATTenuation on page 829
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see " Preamplifier " on page 372.
Remote command:
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe on page 830
Sweep Points
Sets the number of sweep points for the specified range.
For details on sweep points, see Chapter 8.5.1.8, "How Much Data is Measured:
Sweep Points and Sweep Count", on page 381.
Remote command:
[SENSe:]LIST:RANGe<ri>:POINts[:VALue] on page 831
Stop After Sweep
This command configures the sweep behavior.
"On"
The R&S FSV/A stops after one range is swept and continues only if
you confirm (a message box is displayed).
"Off"
The R&S FSV/A sweeps all ranges in one go.
Remote command:
[SENSe:]LIST:RANGe<ri>:BREak on page 827
Transducer
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
The transducer overlaps or equals the span of the range.
The x-axis is linear.
Spurious Emissions Measurement