EasyManua.ls Logo

R&S FSV3000 Series - Page 830

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
830User Manual 1178.8520.02 ─ 01
Manual operation: See " RF Attenuation Mode " on page 258
[SENSe:]LIST:RANGe<ri>:INPut:GAIN:STATe <State>
This command turns the preamplifier for a spurious emission measurement range on
and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Preamp " on page 258
[SENSe:]LIST:RANGe<ri>:INPut:GAIN[:VALue]
<Gain>
This command selects the preamplification level for the range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
[SENSe:]LIST:RANGe<ri>:LIMit:STARt <Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level> Absolute limit at the start frequency of a SEM range.
Range: -400 to 400
*RST: 13
Default unit: dBm
Manual operation: See " Abs Limit Start / Abs Limit Stop " on page 259
[SENSe:]LIST:RANGe<ri>:LIMit:STATe <State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Manual operation: See " Limit Check " on page 259