Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
259User Manual 1178.8520.02 ─ 01
●
The unit is dB.
For details on transducers, see Chapter 12.5.1, "Basics on Transducer Factors",
on page 592.
Remote command:
[SENSe:]LIST:RANGe<ri>:TRANsducer on page 832
Limit Check
Activates or deactivates the limit check for all ranges.
For details on limit checks, see Chapter 7.7.3.2, "Limit Lines in Spurious Measure-
ments", on page 254.
"ABSOLUTE"
Signal is checked against absolute limit values
"NONE"
No limit check is performed.
Remote command:
[SENSe:]LIST:RANGe<ri>:LIMit:STATe on page 830
CALCulate<n>:LIMit<li>:FAIL? on page 1055
Abs Limit Start / Abs Limit Stop
Sets an absolute limit value at the start or stop frequency of the range [dBm].
Remote command:
[SENSe:]LIST:RANGe<ri>:LIMit:STARt on page 830
[SENSe:]LIST:RANGe<ri>:LIMit:STOP on page 831
Insert before Range / Insert after Range
Inserts a new range to the left of the currently focused range (before) or to the right
(after). The range numbers of the currently focused range and all higher ranges are
increased accordingly. The maximum number of ranges is 30.
Delete Range
Deletes the currently focused range. The range numbers are updated accordingly.
7.7.4.2 Adjusting the X-Axis to the Range Definitions
Access: "Overview" > "Select Measurement" > "Spurious Emissions" > "Adjust X-Axis"
The frequency axis of the measurement diagram can be adjusted automatically so that
the span of all sweep list ranges corresponds to the displayed span. Thus, the x-axis
range is set from the start frequency of the first sweep range to the stop frequency of
the last sweep range.
Remote command:
[SENSe:]LIST:XADJust on page 835
7.7.4.3 List Evaluation
Access: "Overview" > "Select Measurement" > "Spurious Emissions" > "List Evalua-
tion"
Spurious Emissions Measurement