Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
835User Manual 1178.8520.02 ─ 01
14.5.7.4 Adjusting the X-Axis to the Range Definitions
[SENSe:]LIST:XADJust
Sets the x-axis range for the spurious emission measurement from the start frequency
of the first sweep range to the stop frequency of the last sweep range.
Example:
SENS:LIST:XADJ
Usage: Event
14.5.7.5
Performing a Spurious Measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST, see [SENSe:]SWEep:MODE on page 793
INITiate<n>[:IMMediate] on page 747, see Chapter 14.5.1, "Performing Mea-
surements", on page 745
14.5.7.6 Retrieving and Saving Settings and Results
The following commands analyze and retrieve measurement results for Spurious mea-
surements.
Useful commands for spurious emission measurements described elsewhere
●
CALCulate<n>:LIMit<li>:FAIL? on page 1055
●
TRACe<n>[:DATA] on page 975
●
TRACe<n>[:DATA]:MEMory? on page 976
●
TRACe<n>[:DATA]:X? on page 976
14.5.7.7 Programming Example: Spurious Emissions Measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Configuring and Performing Measurements