Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
793User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:PRESet:STORe <Standard>
This command saves the current SEM measurement configuration.
Standard definitions are stored in an xml file. The default directory for SEM standards
is C:\R_S\INSTR\\sem_std.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Standard> String containing the file name.
You can save the file in a subdirectory of the directory men-
tioned above. In that case, you have to include the relative path
to the file.
Manual operation: See " Save Standard " on page 235
14.5.6.2 Controlling the Measurement
The following commands control the measurement itself.
INITiate<n>:ESPectrum.................................................................................................. 793
[SENSe:]SWEep:MODE................................................................................................. 793
INITiate<n>:ESPectrum
This command initiates a Spectrum Emission Mask measurement.
Suffix:
<n>
.
irrelevant
[SENSe:]SWEep:MODE <Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
You can select other measurements with
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
Parameters:
<Mode> LIST | AUTO | ESPectrum
AUTO
Turns on basic spectrum measurements.
ESPectrum
Turns on spectrum emission mask measurements.
LIST
Turns on spurious emission measurements.
*RST: AUTO
Configuring and Performing Measurements