Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
750User Manual 1178.8520.02 ─ 01
OBANdwidth | OBWidth
Occupied bandwidth.
Returns the occupied bandwidth in Hz.
Manual operation: See " C/N " on page 198
See " C/N0 " on page 198
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:SELect <MeasType>
This command selects a power measurement and turns the measurement on.
Suffix:
<n>
.
Window
<m> Marker
<sb> irrelevant
Parameters:
<MeasType> ACPower | MCACpower
Adjacent channel leakage ratio (ACLR), also known as adjacent
channel power or multicarrier adjacent channel.
The R&S FSV/A performs the measurement on the trace
selected with [SENSe:]POWer:TRACe.
CPOWer
Channel power measurement with a single carrier.
The R&S FSV/A performs the measurement on the trace
selected with [SENSe:]POWer:TRACe.
OBANdwidth | OBWidth
Occupied bandwidth measurement.
The R&S FSV/A performs the measurement on the trace that
marker 1 is positioned on.
Manual operation:
See " C/N " on page 198
See " C/N0 " on page 198
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe] <State>
This command turns a power measurement on and off.
Suffix:
<n>
.
irrelevant
<m> irrelevant
<sb> irrelevant
Parameters:
<State> ON | OFF | 1 | 0
ON | 1
The power measurement selected with CALCulate<n>:
MARKer<m>:FUNCtion:POWer<sb>:SELect is activated.
Configuring and Performing Measurements