Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1116User Manual 1178.8520.02 ─ 01
14.10.4.2 Programming Example: Using Touchstone files
This example demonstrates how to use Touchstone files in a basic I/Q Analyzer mea-
surement setup in a remote environment.
-----------Configuring the measurement ------------
*RST; *WAI
SYST:DISP:UPD ON
TRACe1:IQ ON
TRACe:IQ:EVAL ON
LAY:REPL:WIND "1",FREQ
INIT:CONT OFF
// Use calibration signal as input data
DIAG:SERV:INP CAL
DIAG:SERV:INP:RF:SPEC BRO
DIAG:SERV:INP:PULS:CFR 1MHz
// Set SR/Bandwidth and CF
TRAC:IQ:SRAT 100 MHz
FREQ:CENT 500MHz
// Activate S2P file
SENSe:CORRection:FRESponse:Input1:USER:PRESet
SENSe:CORRection:FRESponse:Input1:USER:SLISt1:SELect 'c:\BP_40MHz.s2p'
SENSe:CORRection:FRESponse:Input1:USER:STATe ON
INIT:IMM; *WAI
@LOC
14.10.5 Customizing the Screen Layout
The remote commands required to set up the display of the R&S FSV/A are described
here.
● General Display Settings and Items.................................................................... 1116
● Colors and Themes............................................................................................. 1119
● CMAP Suffix Assignment.................................................................................... 1121
14.10.5.1 General Display Settings and Items
The following commands add, remove or customize general display and screen ele-
ments.
Useful commands for general display settings described elsewhere
●
DISPlay[:WINDow<n>]:MTABle on page 988
●
DISPlay:FORMat on page 875
Configuring the R&S
FSV/A