Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1115User Manual 1178.8520.02 ─ 01
For details, see Chapter 12.6, "Frequency Response Correction (R&S FSV/A-K544)",
on page 604.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example:
SENS:CORR:FRES:USER:STAT
Manual operation: See "State" on page 608
[SENSe:]CORRection:FRESponse:BASeband:USER:STORe <FilePath>
[SENSe:]CORRection:FRESponse:INPut<ip>:USER:STORe <FilePath>
[SENSe:]CORRection:FRESponse:USER:STORe <FilePath>
Stores a saveset for a user-defined frequency response correction scenario. By
default, the saveset is stored in the C:\R_S\INSTR\USER\FResponse directory.
Setting parameters:
<FilePath> string
Example:
SENS:CORR:FRES:USER:STOR 'FRes1'
Stores the current frequency response correction configuration
to a file named
C:\R_S\INSTR\USER\FResponse\FRes1.dfl.
Usage: Setting only
Manual operation: See "Save Settings" on page 609
[SENSe:]CORRection:FRESponse:USER:VALid?
This command queries the validity of the user-defined correction settings.
Return values:
<Validity> 0 | 1
1
The setting is valid.
0
The setting is not valid.
Example:
SENS:CORR:FRES:USER:VAL?
Usage: Query only
Manual operation: See "State" on page 608
Configuring the R&S FSV/A