General Instrument Setup
R&S
®
FSVA3000/ R&S
®
FSV3000
608User Manual 1178.8520.02 ─ 01
Clear Settings..............................................................................................................609
Touchstone File Information........................................................................................609
ID.................................................................................................................. 609
Filename....................................................................................................... 609
Active............................................................................................................ 609
To - From.......................................................................................................609
Add Touchstone File....................................................................................................610
Remove File................................................................................................................610
Move File Up or Down.................................................................................................610
Frequency Response active........................................................................................610
Frequency Response File Information........................................................................ 610
ID.................................................................................................................. 610
Filename....................................................................................................... 611
Magnitude..................................................................................................... 611
Phase............................................................................................................611
Add Freq Resp File......................................................................................................611
Remove Frequency Response File............................................................................. 611
Preview........................................................................................................................611
Selected File................................................................................................. 611
IQ Mode........................................................................................................ 612
Spectrum Mode.............................................................................................613
State
Activates or deactivates the general usage of user-defined frequency response correc-
tion settings. If activated, the data in the active correction files is combined to create a
filter. This filter is applied to the measurement results of subsequent sweeps.
Remote command:
[SENSe:]CORRection:FRESponse:USER:STATe on page 1114
Testing the validity of the correction setting:
[SENSe:]CORRection:FRESponse:USER:VALid? on page 1115
Apply to
Determines which channels the correction settings are applied to.
Note: In MSRA mode, the settings are always applied to all active measurement chan-
nels.
"All"
The frequency response correction settings are applied to all active
measurement channels with the same input type. A new filter is cal-
culated for each measurement channel as soon as you switch to it, or
when a measurement is performed (e.g. by the Sequencer).
"Channel"
The frequency response correction settings are applied to the cur-
rently selected channel only.
Remote command:
[SENSe:]CORRection:FRESponse:USER:SCOPe on page 1108
Load Settings
Loads a stored saveset for a user-defined frequency response correction scenario.
Existing settings in the dialog box are overwritten. The settings apply to the currently
selected channel or all channels, depending on the Apply to setting.
Frequency Response Correction (R&S
FSV/A-K544)