EasyManuals Logo

R&S FSV3000 Series User Manual

R&S FSV3000 Series
1268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #373 background imageLoading...
Page #373 background image
Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
373User Manual 1178.8520.02 ─ 01
Remote command:
INPut<ip>:GAIN:STATe on page 904
INPut<ip>:GAIN[:VALue] on page 904
Noise Cancellation
The R&S FSV/A can correct the results by removing the inherent noise of the analyzer,
which increases the dynamic range.
In this case, a reference measurement of the inherent noise of the analyzer is carried
out. The measured noise power is then subtracted from the power in the channel that
is being analyzed (first active trace only).
The inherent noise of the instrument depends on the selected center frequency, resolu-
tion bandwidth and level setting. Therefore, the correction function is disabled when-
ever one of these parameters is changed. A disable message is displayed on the
screen. To enable the correction function after changing one of these settings, activate
it again. A new reference measurement is carried out.
Noise cancellation is also available in zero span.
RMS
Average
Sample
Positive peak
Remote command:
[SENSe:]POWer:NCORrection on page 900
8.4.3 Scaling the Y-Axis
The individual scaling settings that affect the vertical axis are described here.
Access: "Overview" > "Amplitude" > "Scale" tab
Or: [AMPT] > "Scale Config"
The remote commands required to define these settings are described in Chap-
ter 14.7.3, "Configuring the Vertical Axis (Amplitude, Scaling)", on page 898.
Amplitude and Vertical Axis Configuration

Table of Contents

Other manuals for R&S FSV3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSV3000 Series and is the answer not in the manual?

R&S FSV3000 Series Specifications

General IconGeneral
TOI (Third-Order Intercept)> +15 dBm
Resolution Bandwidth1 Hz to 1 MHz
Display12.1-inch touchscreen
Analysis Bandwidth40 MHz
Video Bandwidth (VBW)1 Hz to 1 MHz
Amplitude accuracy±0.5 dB
InterfaceUSB
Dynamic Range> 100 dB

Related product manuals