Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1156User Manual 1178.8520.02 ─ 01
14.11.6 Controlling the Positive Transition Part
For more information on the negative transition part see Chapter 13.1.7.2, "Structure of
a SCPI Status Register", on page 667.
STATus:OPERation:PTRansition <SumBit>
STATus:QUEStionable:PTRansition <SumBit>
STATus:QUEStionable:ACPLimit:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:EXTended:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:EXTended:INFO:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:FREQuency:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:LIMit<n>:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:LMARgin<n>:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:POWer:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:TEMPerature:PTRansition <SumBit>,<ChannelName>
STATus:QUEStionable:TIME:PTRansition <SumBit>,<ChannelName>
These commands control the Positive TRansition part of a register.
Setting a bit causes a 0 to 1 transition in the corresponding bit of the associated regis-
ter. The transition also writes a 1 into the associated bit of the corresponding EVENt
register.
Suffix:
<n>
.
Window
Parameters:
<SumBit> Range: 0 to 65535
<ChannelName> String containing the name of the channel.
The parameter is optional. If you omit it, the command works for
the currently active channel.
14.12 Commands for Remote Instrument Operation
The following commands are required to shutdown or reboot the R&S FSV/A from a
remote PC.
SYSTem:CLOGging......................................................................................................1156
SYSTem:REBoot..........................................................................................................1157
SYSTem:SHUTdown.................................................................................................... 1157
SYSTem:CLOGging <State>
This command turns logging of remote commands on and off.
Parameters:
<State> ON | OFF | 1 | 0
Commands for Remote Instrument Operation