Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
353User Manual 1178.8520.02 ─ 01
Defines the type of signal available at the "IF" and "Video" output connectors of the
R&S FSV/A.
For restrictions and additional information, see Chapter 8.2.1.3, "IF and Video Signal
Output", on page 311.
IF Output State............................................................................................................353
IF Out Frequency ....................................................................................................... 353
Video Output State......................................................................................................353
IF Output State
Enables or disables output of the measured IF value at the "IF" output connector.
Remote command:
OUTPut<up>:IF:STATe on page 952
IF Out Frequency
Defines or indicates the frequency at which the IF signal level is provided at the "IF"
output connector if IF Output State is enabled.
For more information, see Chapter 8.2.1.3, "IF and Video Signal Output", on page 311.
Remote command:
OUTPut<up>:IF:IFFRequency on page 952
Video Output State
Enables or disables output of the displayed video signal (i.e. the filtered and detected
IF signal) at the "Video" output connector.
Remote command:
OUTPut<up>:VIDeo:STATe on page 953
8.2.7 Audio Output
Access: [Input/Output] > "Output" > "Audio" tab
The R&S FSV/A can provide audio output to special connectors if the R&S FSV3-B5
option is installed.
For details on connectors, refer to the R&S FSV/A Getting Started manual, "Front /
Rear Panel View" chapters.